Lambda Advanced Technology
Products  Fluorescence  F-2700
Fluorescence Spectrophotometer
F-2700

f2700

Fluorescence Spectrophotometer
F-2710

f2710

Fluorescence spectrophotometers available in stand alone or PC controlled versions.
The Hitachi F2700/2710 fluorescence spectrophotometers are high sensitive variable band pass instruments featuring an ozone free high intensity continuous Xe lamp.

The F-2700/F-2710 feature an excellent signal-to-noise ratio (800 S/N RMS, Raman band of water at 5 nm bandwidth). With PC control the scan speed is as high as 12,000nm/min, making it ideal for 3-D scan measurements. A wide array of accessories is available, including automatic sampling, quantum yield measurement, full-range spectral correction, and 100?L micro cell.
  1. Leading Edge Sensitivity (800 or better RMS)
  2. Wide Dynamic Range (with improved zero point adjustment)
  3. Ease of Use / Stand-alone Operation (no PC required)
  4. Optional PC Operation
  5. Automatic Performance Monitoring
ITEM DESCRIPTION
Operation mode
Stand-alone or PC operation
Light source 150W Xenon lamp (self-deozonating lamp house)
Spectrometer Aberration-corrected concave diffraction grating: 900 lines/mm. Blaze wavelengths 300nm EX, 400nm Em
Wavelength measurement range (Ex and Em) 220 nm to 730 nm and zero order(optional 220 to 800 nm)
Wavelength accuracy ±3.0 nm
Wavelength scan speed 60, 300, 1500, 3000, 12000 nm/min (12,000 nm/min only under PC control)
Wavelength drive speed 12,000 nm/min
Response From 0 to 98%: 0.04, 0.08, 0.4 and 2 sec.
Slit width (Ex and Em) 2.5, 5, 10, 20 nm
Photometric range ± 9999
Photometric principle
Monochromatic light monitoring ration calculation
Sensitivity
800 or more (RMS) (bandpass 5nm, response time 2 S)
Bandpass
2.5, 5, 10, 20 nm (both Ex and Em)
Wavelength scan speed
60 – 3,000 nm/min (four steps), 12,000 (when PC control)
Dimensions / Weight
600 (w) x 503 (D) x 343 (H) mm/ approx.41 kg
Power consumption
AC100, 115, 220, 230, 240 V 50/60 Hz 400 VA